General information
Usage
- Transmission geometry, the gamma radiation is going through the sample, the detector on the other side is collecting absorption spectrum
- Allows to study the sample under low temperatures
Sample types
- Powder
- Thin bulk samples (must be transparent for 14.4 keV gamma photon radiation)
Technical Specifications
Transmission Detector
- YAP:Ce scintillation detector (1“ photomultiplier)
- High Voltage supply built-in (electronically adjustable)
- Pre-amplifier & amplifier built-in (electronically adjustable)
- Single channel analyser built-in (electronically adjustable)
- Detection range 10keV % 30keV
- Spectrum registration built-in (1024 channels per unfolded spectrum)
- Nickel plated brass aperture collimator (hole diameters: 4mm, 6mm, 8mm, 10mm, 14mm)
- Standard USB interface
- PC control application
- Power requirements: +5V/500mA, +15V/20mA, -5V/20mA
Velocity Drive
- Neodymium magnets based electrodynamic linear motor
- Drive coil & sensor coil
- Fast analogue PID velocity controller built-in
- Velocity generator built-in (up to 50 mm/s)
- M4 mounting screw for source
- Internal lead shielding
- Nickel plated brass aperture collimator (hole diameters: 4mm, 6mm, 8mm, 10mm, 14mm)
- Standard USB interface
- PC control application
- Power requirements: +5V/250mA, +15V/100mA, -5V/20mA, -15V/100m
Sample Exchanger
- 2 removable stacks, 7 samples per stack
- Standard USB interface
- PC control application
- Power requirements: +5V/250mA, +15V/300mA
Power Supply
- Standard commerical laboratory power supply
- Typically 4-channel GW Intek GPE-4323 (110V/240V)
- Supply voltages: +5V, +15V, -5V, -15V
Experimental setup
- Standard optical rail and carriages
Calibration set
- MRA.2.6 standard alpha iron reference absorber (ritverc.com)
- Dual-channel oscilloscope (USB)
Control PC
- All-in-one Personal Computer with keyboard and mouse
(23.8″ Full HD, 8GB RAM, 256 SSD, multi-core processor) - Windows 11 operating system
- Remote access
- Interface for controlling the spectrometer (Spectrometer is independent on PC)
General notes
- Turn-key solution, all necessary parts and tools included
- All standard safety standards met
- Internal lead shielding
- Working with both 110V and 240V power grid
- Most of the parameters are customizable
- A radioactive source is necessary for Mössbauer Spectrometer operation (a delivery of the source is possible on customer’s request)
Do you have any questions?